Retsch AS 200 control - natural frequency User Manual

Page 16

Advertising
background image

Retsch GmbH

16

Adjusting the time

12

13

14

On switching on the AS 200 control the sieving time is
undetermined = 2 bars in display 13.

Switching on the time 1 – 99 min:

Button 12 reduces the time down to 1 min. If the value drops

below this, two bars appear = undetermined sieving time.

Button 14 increases the time up to 99 min. If this value is

exceeded two bars appear = undetermined sieving time.

Memory –

storage and calling sieving parameters

2

5

8

10 11

13

The memory function, i.e. storage and calling preselected sieving
parameters, can be carried out only in the standby mode. You can
change and overwrite the programme positions P1 to P9 at any time.

On switching on the AS 200 control the indication „on“ appears in
the display 2.
By pressing the button PROG you can go to the next programme
position P1 to P9. Following programme position P9 „on“ reappears
in the display. In the PROG mode all buttons except PROG, START
and STOP
are blocked.

When on is displayed
it is possible to freely adjust the sieving parameters - 5
amplitude / 8 interval time / 10-11 interval on-off and 13
time.

When P1 to P9 is displayed
Sieving parameters can be stored and called.

Start = the sieving process is started with the selected
programme or the freely selected parameters under “on”.

Allocation of memory location

Press button PROG once or several times until the required
memory location P1-P9 has been reached.

Press button SET – all displays flash.

Set sieving parameters - 5 amplitude / 8 interval time / 10-
11
interval on-off and 13 time.

The programming mode can be aborted by pressing the PROG button
– the values are not stored.

Press SET button – the values are stored

Flashing of the display stops, the adjusting interlock is activated, the
sieving parameters are stored.

If the AS 200 control is driven by a PC and the “EasySieve

®

sieving software the indication “ES” appears in the display 2 . In this
mode no manual changes in the sieving parameters are possible.

Advertising