Working instructions, Sieving aids sieving material quantities, Sieving aids – Retsch AS 200 control - natural frequency User Manual

Page 25

Advertising
background image

Retsch GmbH

25

Working instructions

Sieving aids

When dealing with material that is difficult to separate, we
recommend the additional use of sieving aids in the individual
sieve fractions. Depending on the width of the sieve meshing and
the preselected vibration intensity, pearls made of agate, rubber,
porcelain or nylon bristles and Vulkollan cubes may be used. Also
refer to the table below.

Ensure that the fabric of the sieve is not overstretched by
overloading with sieving aids as this would have a detrimental
effect on the precision of your analytical sieve.

Overview table

Sieving aids

Quantity per

sieve or charge

Supplier

Area of use

Caution !!

Rubber pearls

5 pcs, Ø20 mm

RETSCH, Haan

Materials to be screened
are fine, dry,
difficult to separate and
throw sieves are used

When using mechanical sieving aids, there
is the danger that soft materials can
disintegrate and fine sieve fabrics can be
damaged!

Agate pearls

10 pcs, Ø10 mm

RETSCH, Haan

Hard porcelain pearls

approx. 10gr., Ø2
mm

RETSCH, Haan

Vulkollan cubes

5 pcs, 12x12

RETSCH, Haan

Planetary sieving
machines

Bristles

3 pcs

RETSCH, Haan

Plastic rings

3 pcs, various Ø

Brushes

Manual sieving

Highly dispersed silicic acids

0.5 to 2%

Degussa, Ffm

Materials to be screened
are fine, sticky, fatty
and electrostatically
charged

Check applicability and compatibility before
adding!

- - Aluminium oxide

0.5 to 1%

Degussa, Ffm

- - active carbon

- - talcum

Latex and rubber
powder

Remove grease

Materials to be screened
are very fatty

The granularity of the material to be
screened may not be altered by the solvent
used to remove grease or by the
temperature and air used in drying.

Dry

Materials to be screened
are very wet or damp

Guide dry, warm air through the
sieve set

Hygroscopic materials to
be screened

Cleanse with steam

Materials to be screened
are coarse and
electrostatically charged

First sieve of the fine sieve then
the coarse sieve

If contamination from
other mechanical aids,
such as rubber pearls, is
to be avoided

The coarse material functions like a
mechanical sieving aid (pearls) on the fine
sieves.

Intermittent sieving

RETSCH, Haan

Materials to be screened
are fine, stick easily, or
only fall through the
mesh with difficulty

Intermittent switching is built into the
majority of throw sieves and can be
switched on, if required.

Sieving material quantities
The sieve stack required for sieve analyses is composed of
analytical sieves arranged one above the other, with
progressively larger holes towards the top, plus the base pan.

To ensure fast separation with exact results, the quantity of the
material to be screened should be adapted to the sieve diameter
and the nominal size of the openings.
The relevant standards on sieve analysis state guideline values in
this context.

In the case of sieving processes with only 1 to 3 sieves, we
recommend the use of short threaded rods ( available as
accessories ).
Overly long, protruding rods interfere with the material to be
screened due to their inherent vibration characteristics.