Servicing the colonydoc-it, Cleaning and care – UVP ColonyDoc-It Imaging Station User Manual

Page 45

Advertising
background image

ColonyDoc-It Imaging Station

45

Print an Audit Trail (Image History)

Open the image for which an Audit

Trail print copy is needed.

Go to

Tools > Reports. (This option

is disabled, if no printer is available.)

A window opens with various types

of reports available.

To select to print an Audit Trail, click

the

Image History item. If other

reports are required, click in the

checkbox of the other reports

needed. For example, if both an

audit trail and the image report need

to be printed, click on

Image Report

and

Image History.

Adjust the header and footer settings

or printer settings if necessary, and

print the trail.


Servicing the ColonyDoc-It

Cleaning and Care

Use 70% Isopropanol Alcohol to clean the ColonyDoc-It external surfaces, base, and doors.
Clean the ColonyDoc-It twice a day, before and after every use.

Cleaning the Diopter

Weekly clean the diopter with a soft photographic lens cleaning cloth dampened with lens cleaning solution.

CAUTION: Do not pour the solution directly onto the diopter.
The diopter is located in the filter box. To access the diopter:
Remove the four brass

thumbnuts on the filter box

and remove the filter box

from the housing.

Clean both sides of the

diopter. Remove the filter if

necessary to access the

diopter (refer to Installing
the Filter

for instructions on

removing the filter.)