3.1 General rules

748 DH Sample Changer

11

3 Operation via RS232 Interface

3.1 General rules

The DH Sample Changer has a remote control facility that allows its control
via TiNet. The parameters of the Sample Changer may be easily set by the
method editor of TiNet.
The following description offers special possibilities of low level control via
the transfer command of the calculation block of the TiNet method editor.

3.1.1 Tree of commands

The commands are hierarchically ordered with the following rules:

Rules

Example

The root of the tree is designated by &.

The branches (levels) of a tree are marked with

a dot (.) when calling up an object.
Upper- or lowercase letters may be used.

&STIRER.ON or &Stirer.On

An object can be assigned a value. Values are

marked at the beginning and end by quotes (").

They may contain ASCII characters or numeri-

cal values.

&Stirer.On"True"

&Init.SpeedXY"3"

The current object remains valid until a new

object is called.
New objects can be addressed relative to the

old object:

A preceding dot leads forwards to the next level

in the tree.

More than one preceding dot leads one (or

several) level(s) backwards in the tree. n node

backwards require n+1 preceding dots.

&Stirer.On"True";..$G

Leads back from nod &Stirer.On to nod &Stirer

If you want to jump back to the root, enter a

preceding &.

3.1.2 Triggers

$G

Go

Commands are executed with $G.

Example:
&Stirer.On"True"

Sets command.

..$G

Starts stirrer.