4.2 TiNet error messages

748 DH Sample Changer

18

4.2 TiNet error messages

Problem

Possible causes and remedies

Invalid head position selected.

A rack position has been selected which is not

contained in the rack definition file.

Check the selected position.

Have you selected the correct rack in the

TiNet configuration?

Selected command can't be executed.

General communication problem. Exchange

the cable.

There is no beaker. (Position: X)

In the TiNet configuration the beaker test is set

to "Display" and no beaker is present in the

given position.

Set the beaker test to "Move" if you want to

move to the next beaker automatically or to

"OFF" if you want to switch off the beaker test.

There is no reply from the Sample

Changer.

Communication interrupted?

Try to initialize the Sample Changer with the

menu item "Commands, Initialize

instruments" in the titration part of TiNet.

Reboot computer.

Switch Sample Changer off/on, make sure

cable is well plugged in and try to initialize

again.

Does the green lamp light up? If not, check

whether the mains cable is plugged in

correctly. If everything seems to be OK,

change the power supply fuses, see page 5.

If the green lamp lights up and everything

appears to be OK then change the pump

fuse, see page 5.

Exchange cable.

Transfer error: the target device cannot

receive data.

Errors if transfer commands from the

calculation block in the method editor are

used:

Wrong path, value and/or trigger.

Communication interrupted.