3 readjust sample changer – Metrohm 748 DH Sample Changer User Manual

Page 21

Advertising
background image

4.3 Readjust Sample Changer

748 DH Sample Changer

19

4.3 Readjust Sample Changer

1. Load the test software from the Internet:

http://www.metrohm.ch

2. Remove the sample racks from the Sample Changer.
3. Install 6.2625.000 position guide on the sample arm in-

stead of the titration head.

4. Place the centering ring over the right-hand side rack

bolt.

5. Start the test software which you have loaded from the

Internet and follow the instructions which you will also
find there.

Sample arm from below
with position guide:

Centering ring over rack
bolt: