5.1 Technical specifications

748 DH Sample Changer

20

5 Appendix

5.1 Technical specifications

Lift

Max. lift way

146 mm

Max. load

500 g

Sample position

precision

±

2 mm

Sample racks

48 beaker, 250 mL
136 beaker, 75 mL

Material

Light metal, powder coated

Pump

Capacity

0.33 L/min

Fuse

for 100...120 V: T1.25A
for 220...240 V: T630mA L

Housing

Al (antichromed) and stainless steel, powder coated

RS232 interface

for PC connection

Input/output lines

for beaker sensor and any other devices

Ambient temperature

Nom. operation range

0 ... 45 °C

Storage, transport

– 25 ... 60 °C

Safety specifications

Designed and tested in accordance to IEC publication
1010-1, safety class I. This manual contains information
and warnings which have to be followed by the user to en-
sure safe operation and to retain the apparatus in safe
condition.

Mains connection

Voltage

100, 120, 220, 240 V

±

10 % (switchable)

Frequency

50 ... 60 Hz

Power consumption

max. 190 VA

Fuses (2)

for 100...120 V: T2A/120V
for 220...240 V: T1A/250V

Dimensions

Width

750 mm

Height

750 mm

Depth

600 mm

Weight, incl. accessories

ca. 35.5kg