3.3 Mounting the extension module below the IC instrument

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

872 Extension Module IC Pump

9 Mounting the drainage tubing

Reconnect the drainage tubing to the drainage tubing connector of
the bottle holder.

Possibly, a longer section of silicone tubing 6.186.020 must be cut to
fit and mounted (see also the manual for the IC instrument).

3.3

Mounting the extension module below the IC instru-
ment

1 Switching off the IC instrument

Switch the IC instrument off and disconnect the mains cable.

2 Clearing the bottle holder

If there are bottles and other things on the bottle holder, remove
them.

3 Disconnecting all connections on the rear of the instrument

Disconnect the mains cable,

Disconnect the MSB cable,

Disconnect the USB cable,

Disconnect the leak sensor,

Remove the drainage tubings.

4 Removing the detector(s)

Disconnect the detector cable(s) and remove the detector(s) from the
IC instrument (see manual for the IC instrument).

5 Removing the base tray

Tilt the IC instrument sideways and lay it down flat.

Loosen the cylinder screws with a 6.2621.100 3 mm hexagon
key.

Remove base tray.