7.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

54

■■■■■■■■

872 Extension Module IC Pump

7 Troubleshooting

7.1

Problems and their solutions

Problem

Cause

Remedy

Marked rise in pres-
sure

6.2821.120 inline filter
blocked.

Replace 6.2821.130 filter (see Chapter 6.5,
page 51)
.

Drift of the baseline

Eluent – Evaporation of the
organic solvent in the elu-
ent.

Check the eluent bottle attachment (see Figure
16, page 26)
.

Very noisy baseline

High pressure pump – con-
taminated pump valves.

Clean pump valves (see Chapter 6.4.2, page
41)
.

Eluent – Leak in the eluent
path.

Check the eluent path.

Eluent – Blockage in the
eluent path.

Check the eluent path.

High pressure pump –
defective piston seals.

Replace piston seals (see Chapter 6.4.2, page
41)
.

Pulsation damper not con-
nected.

Connect pulsation damper (see Chapter 4.8,
page 35)
.

Background conduc-
tivity too high

Incorrect eluent.

Change eluent (see Chapter 6.3.2.3, page 40).

Poor reproducibility
of the retention
times

Eluent – Leak in the eluent
path.

Check the eluent path.

Eluent – Blockage in the
eluent path.

Check the eluent path.

Unexpected change
to the retention
times in the chroma-
tograms

Eluent – Gas bubbles in the
eluent.

Check connections of the eluent degasser (see
Chapter 4.5, page 28)
.

High pressure pump –
defective.

Request Metrohm Service.

Vacuum is not being
built

Eluent Degasser – Connec-
tor Vacuum on the rear of
the instrument not (tightly)
sealed.

Seal the connector Vacuum tightly with a
6.144.6040 threaded stopper.
or when using an 872 Extension Module: