4.5 Eluent degasser

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

872 Extension Module IC Pump

5

Eluent

6

Aspiration filter 6.2821.090

7

Filter holder
From accessory set 6.2744.210.

8

Clamping screw
From accessory set 6.2744.210.

9

Tubing weighting
From accessory set 6.2744.210.

10 Eluent bottle 6.1608.070

11 Bottle attachment 6.1602.160

12 SGJ clip 6.2023.020

13 Tubing nipple

14 Thread stopper

4.5

Eluent degasser

Gas bubbles in the eluent lead to an unstable baseline, as high pressure
pumps can transport liquids, but not gases. The eluent therefore has to be
degassed, before it reaches the high pressure pump.

The eluent degasser removes gas bubbles and dissolved gases from the
eluent. For degassing, the eluent flows into a vacuum chamber through a
special fluoropolymer capillary.

Note

The eluent degasser is already installed in the newly delivered instru-
ment. The following installation instructions need not be carried out at
the time of initial installation.

Connecting the eluent degasser

5

6

1

2

3

4

Figure 19

Eluent degasser

1

Eluent degasser input

2

Eluent degasser output