Figure 11, Connecting the vacuum pump – Metrohm 872 Extension Module IC Pump User Manual

Page 28

Advertising
background image

3.5 Connecting the vacuum pump(s)

■■■■■■■■■■■■■■■■■■■■■■

20

■■■■■■■■

872 Extension Module IC Pump

“Vacuum”
850 Professional IC

“to Vacuum”

872 Extension Module 1

“to Vacuum”
872 Extension Module 2

1

2

4

3

Figure 11

Connecting the vacuum pump

1

T connector 6.1808.060
For connecting several FEP tubings.

2

FEP tubing 6.1805.050
Connect to the connector to Vacuum on the
first 872 Extension Module.

3

FEP tubing 6.1805.050
Connect to the connector to Vacuum on the
second 872 Extension Module.

4

FEP tubing 6.1805.060
Connect with the connector Vacuum on the
850 Professional IC instrument.

1 Connect one end of the 6.1805.050 FEP tubing (11-2) to the con-

nector to Vacuum of the 872 Extension ModuleIC Pump.

Connect the other end of the FEP tubing to the 6.1806.060 T con-
nector (11-1).

2 Connect one end of the second 6.1805.050 FEP tubing (11-3) to the

connector to Vacuum of the second 872 Extension Module IC
Pump.

Also connect the other end of the FEP tubing to the 6.1808.060 T
connector.

3 Connect one end of the 6.1805.060 FEP tubing (11-4) to the third

output of the 6.1808.060 T connector.

Connect the other end of the FEP tubing to the connector Vacuum
of the 850 Professional IC instrument.