4 installation, 1 about this chapter, 2 installation overview – Metrohm 872 Extension Module IC Pump User Manual

Page 29: About this chapter, Installation overview

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Installation

872 Extension Module IC Pump

■■■■■■■■

21

4 Installation

4.1

About this chapter

The Installation chapter contains

this overview

a brief set of instructions for the installation of the 872 Extension Mod-
ule IC Pump (see Chapter 4.2, page 21). At each step you will find
cross-references to detailed installation instructions for individual com-
ponents, should you require such aids.

two installation diagrams (see Chapter 4.3, page 22), which show
two possible applications of the instrument.

several chapters with detailed installation instructions for all compo-
nents, including those that are already installed at the time the instru-
ment is delivered.

4.2

Installation overview

Note

A number of the capillary connections are already connected at the
time the instrument is delivered.

The following work steps must still be carried out:

Installing 872 Extension Module IC Pump

1 Installing the eluent path

Equip the eluent aspiration tubing (12-1)/(13-1) and connect with
the eluent bottle (see Chapter 4.4, page 24).

Connect the eluent degasser input (see Chapter 4.5, page 28).

Shorten the 6.1831.100 connection capillary (12-2)/(13-2) to the
required length using the 6.2621.080 capillary cutter and connect
it to the pulsation damper with a 6.2744.070 pressure screw
(12-3)/(13-3) (see Chapter 4.8, page 35).