5 Operation and maintenance

850 Professional IC

■■■■■■■■

95

1

2

3

4

5

Figure 47

MSM-HC – Components

1

Union nut

2

MSM-HC connecting piece 6.2835.010

3

MSM-HC rotor A 6.2842.000

4

MSM-HC housing

5

Slot in the MSM-HC housing

Cleaning the MSM-HC

Clean the MSM-HC as follows:

1 Disconnecting the MSM-HC from the IC system

Switch off the instrument.

Disconnect the MSM-HC from the separation column, peristaltic
pump and detector.

2 Dismantling the MSM-HC

Unscrew union nut (47-1) from the MSM-HC housing (47-4).

Pull the MSM-HC connecting piece (47-2) and the MSM-HC rotor
A (47-3) out of the MSM-HC housing (47-4). The connecting
piece and the MSM-HC rotor A normally stick to one another – if
this is not the case: Take a sharp object, insert into the slot (47-5)
in the MSM-HC housing, and pull out the MSM-HC rotor A (47-3)
in this way.

Detach the MSM-HC connecting piece (47-2) from MSM-HC rotor
A (47-3).

3 Cleaning the supply and discharge lines

Connect in turn each of the 6 capillary tubings fastened on the
MSM-HC connecting piece (47-2) on the high pressure pump (see
Chapter 3.10, page 37)
and pump through ultra pure water.