Metrohm 850 Professional IC Anion MSM-HC MCS Gradient User Manual

Page 24

Advertising
background image

3.2 Initial installation

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

850 Professional IC

Connect drainage tubings (see "Installing drainage tubing", page
29)
.

3 Connecting the eluent path

Equip an eluent aspiration tubing for each of the two high pres-
sure pumps and connect with the eluent bottle (see Chapter
3.8.1, page 32)
.

Instead of the column, connect the 6.2744.040 coupling with a
6.2744.010 pressure screw at the end of the pre-installed column
inlet capillary.

Use a 6.2744.014 pressure screw to connect the capillary of the
MSM-HC labeled with Eluent to the other end of the 6.2744.040
coupling.

Use a 6.2744.090 long pressure screw to connect the capillary of
the MSM-HC labeled with Detector to the input of the MCS (see
Chapter 3.18.2, page 60)
.

Connect the detector inlet capillary with a 6.2744.090 long pres-
sure screw to the output of the MCS (see Chapter 3.18.2, page
60)
.

4 Connecting the sample path

Note

The sample degasser does not have to be connected. We recom-
mend the usage of the sample degasser only if the sample matrix
requires it.

Connect the 6.1803.040 sample aspirating capillary connected at
the sample input of the injection valve to the output of the sam-
ple degasser with the aid of a 6.2744.090 long pressure screw
(see Chapter 3.13, page 45).

Connect a part of the 6.1803.040 PTFE capillary with a
6.2744.090 long pressure screw to the input of the sample
degasser. Guide the other end out of the instrument through a
capillary feed-through and connect it to the Sample Processor.

Guide the sample outlet capillary connected to the sample output
of the injection valve out of the instrument through a capillary
feed-through and onward to the waste container and then fasten
it there.

5 Installing the peristaltic pump

(see Chapter 3.17.2, page 56).

Advertising