3 Installation

850 Professional IC

■■■■■■■■

15

Connect the 6.1803.020 aspirating capillary for regeneration solu-
tions with a 6.2744.034 tubing olive and a 6.2744.070 short
pressure screw to the aspiration end of the 6.1826.320 pump
tubing for the regeneration solution.

Place the pump tubing in a tubing cartridge.

Connect the 6.1803.020 aspirating capillary for rinsing solutions
with a 6.2744.034 tubing olive and a 6.2744.070 short pressure
screw to the aspiration end of the second 6.1826.320 pump tub-
ing for the rinsing solution.

Place the second pump tubing in the second tubing cartridge.

Place both tubing cartridges into the peristaltic pump.

6 Connecting the MSM-HC

(see Chapter 3.16.2, page 52)

Connect the capillary labeled with H2SO4 with a 6.2744.180
pump tubing connection and a 6.2744.070 short pressure screw
to the peristaltic pump on the outlet end of the pump tubing for
the regeneration solution.

Connect the capillary labeled with H20 with a 6.2744.180 pump
tubing connection and a 6.2744.070 short pressure screw to the
peristaltic pump on the outlet end of the pump tubing for the
rinsing solution.

Guide the two capillaries labeled with Waste into a waste con-
tainer and fasten them there.

7 Connecting the instrument

Connecting the instrument to the PC with a 6.2151.020 USB
cable (see Chapter 3.20.1, page 66).

Connect instrument to mains supply (see Chapter 3.20, page
66)
.

8 Initial start-up

(see Chapter 4.1, page 71)

Switch on the PC and start MagIC Net.

Switch on the instrument.

Deaerate the high pressure pump (see Chapter 3.10.2, page
40)
.

Set contact pressure of the peristaltic pump (see "Set flow rate",
page 59)
.

Rinse instrument without column.