2 Installation

812 Valve Unit

8

2.2 Mains

connection

The 812 Valve Unit is operated with the 6.2152.000 Power supply unit

which automatically adjusts itself to the existing mains voltage

(100

240 V) and frequency (50

60 Hz).

The power supply unit is connected to the mains using one of the fol-

lowing mains cables

6.2122.020 with plug SEV 12 (Switzerland,

)

6.2122.040 with plug CEE(7), VII (Germany,

)

6.2133.070 with plug NEMA 5-15 (USA,

)

which are three-cored and fitted with a plug with an earthing pin.

The cable permanently mounted to the power supply unit is plugged

into connection 7 at the 812 Valve Unit (see Fig. 2). The operational

readiness is shown by the LEDs lighting up on the "FILL" or "INJECT"

keys.

2.3 Electrical

connection

2.3.1

Connection at the 762 IC Interface

Always switch off 762 IC Interface and 812 Valve Unit before you

connect the two instruments.

The connection of the 812 Valve Unit at the 762 IC Interface enables the

three following functions:

Control of valve A by the 762 IC Interface

via remote line

Control of valve B by the 762 IC Interface

via remote line

Automatic start of data acquisition at the

762 IC Interface by switching valve A or valve B

to the "INJECT" position

For instrument connection proceed as follows (see Fig. 3):

1 Connection for control of valve A

Connect the plugs inscribed with "GND", "INJECT" and "FILL"

of the 6.2128.100 cable to the connections "Ground", "Fill"

and "Inject" on the terminal block 10 of the 812 Valve Unit

(see Fig. 2).

Connect the plug inscribed with "Inject NO" of the 6.2128.100

cable to the desired remote line 1...7 "Events - Run" at the

762 IC Interface and the plug inscribed with "Inject COM" to

the same remote line "Events - COM" (see Fig. 5 Instructions

for use 762).