Metrohm 812 Valve Unit User Manual

Page 18

Advertising
background image

3.2

Operation via «IC Net»

812 Valve Unit

15

Valve A

<Fill>

Switch valve A to "FILL" position.

<Inject>

Switch valve A to "INJECT" position.

Valve B

<Fill>

Switch valve B to "FILL" position.

<Inject>

Switch valve B to "INJECT" position.



Time program

On the

Program

tab of the

812 Valve Unit

window a user-defined time

program can be entered. This program starts automatically as defined

in the

Start mode

window either at the moment the determination is star-

ted (

Start with determination

) or at the moment the sample is injected

(

Start with inject

).

Time (1

st

column)

Time at which program instruction is applied.

Range:

0.0 ... 999.9 min

If no time is entered, the program instruction is

applied together with the last instruction with

time entry.

Instruction (2

nd

column)

Program

instruction (see below).

Parameter (3

rd

column)

Parameter for program instruction (see below).

ENABLED

Enable program start (a disabled program is

not started).

<

Add

>

Add new program instruction.

<

Delete

>

Delete selected program instruction.

<

Verify

>

Test the time program (error messages are dis-

played if program is wrong).