Metrohm 812 Valve Unit User Manual

Page 14

Advertising
background image

2.4 Connections at injection valve

812 Valve Unit

11

2

1

4

5

3

6

13

14

15

17

16

Fig. 5: Connection at injection valve

13

Syringe tubing

6.1803.020 PTFE tubing with

6.2744.120 coupling and

6.2816.020 syringe

16

Eluent outlet

6.1831.010 PEEK

capillary

14

Aspirating tubing

6.1803.020 PTFE tubing for

aspirating the sample

17

Eluent inlet

6.1831.010 PEEK

capillary

15

Sample loop

6.1825.210 PEEK sample

loop (20

µ

L )

2 Connecting the aspirating tubing

Cut off a piece of the 6.1803.020 PTFE capillary to the de-

sired length (at best using the optionally available 6.2621.080

tubing cutter) and fit it on one side with a 6.2744.010 PEEK

compression fitting (do not tighten too firmly).

Screw this end of the aspirating tubing 14 to connection 2 of

the injection valve.

3 Connecting the sample loop

Fit the 6.1825.210 sample loop on both sides with a

6.2744.010 compression fitting.

Screw the sample loop 15 (20

µ

L) to connections 3 and 6 of

the injection valve.

4 Connecting the eluent supply

Screw the PEEK capillary 17 for eluent supply from the high

pressure pump to connection 5 of the injection valve.

5 Connecting the separating column

Screw the PEEK capillary 16 for eluent supply to the separat-

ing column to connection 4 of the injection valve.