Figure 3, Setting the mains voltage – Metrohm 816 IC Eluent Selector User Manual

Page 14

Advertising
background image

2 Installation

1 Pull out mains cable

Remove the mains cable from mains supply connection 9 of the

816 IC Eluent Selector.

2 Remove the fuse holder

Use a screwdriver to loosen fuse holder 12 below the mains

supply connection and remove it completely.

3 Check the fuse

Carefully remove the built-in fuse for the intended voltage from the

fuse holder and check its specifications (the position of the fuse in

the fuse holder is indicated by the white arrow beside the voltage

range):

100

120 V 1.0 A (slow blow) Metrohm No. U.600.0016

220

240 V 0.5 A (slow blow) Metrohm No. U.600.0013

4 Insert fuse

Exchange the fuse if necessary and replace it in the fuse holder.

5 Insert fuse holder

Depending on the required mains voltage, insert the fuse holder

in the 816 IC Eluent Selector so that the correct voltage range can

be read normally and the white arrow beside it points to the white

bar printed below the fuse holder (see Figure 3).

220 – 240 V

100 – 120 V

8 Mains switch

9 Mains connec-

tion

12 Fuse holder

220

- 240 V

100 - 1

20

V

220 - 2

40

V

100

- 120 V

12

9

8

Figure 3:

Setting the mains voltage

816 IC Eluent Selector/ 8.816.1003 Instructions for Use

8

Advertising