2 Installation

1 Attaching the eluent inlet tubing

Screw the PEEK fittings of the three pieces of 6.1834.020

Aspiration tubing supplied loosely by hand into connec-
tions "A" and "B" of Valve I 5 and "C" of Valve II 7 of the IC

Eluent Selector. Tighten the screw fittings by using the

6.2621.050 ¼" Wrench (also included in the standard

equipment) to rotate them a further ¼ turn.

Make sure that the other ends of the aspiration tubing are

immersed in the eluents.

2 Attaching the outlet to the IC pump

Screw the 6.1834.030 Tubing connection 816-709 by hand
into the "PUMP" connection of Mixing chamber 6. Tighten

up the connection by using the 6.2621.050 ¼" Wrench to

rotate it a further ¼ turn.

Slide the tubing end onto the capillary at the aspiration con-

nection of the IC pump.

Make sure that the tubing between the outlet of the IC Eluent Selector

and the IC pump is kept as short as possible by shortening the

6.1834.030 Tubing connection 816-709 accordingly. In this way you

minimize the dead volume and therefore the reaction time after

switching to a different eluent.


2.5 Software

installation

The PC program «IC Net 2.1» is required for controlling the 816 IC Elu-

ent Selector from a PC. This program runs under the operating systems

Windows 95, Windows 98, Windows NT and Windows 2000 and is in-

stalled as described in Section 1.4.2 of the 8.110.8223 Instructions

for Use for the Metrodata «IC Net 2.1» software.
The installation of the 816 IC Eluent Selector is described in Section

6.12 of the 8.110.8223 Instructions for Use for the Metrodata «IC Net

2.1» software.
Make sure that the RS 232 settings of the 816 IC Eluent Selector and

the interface of the connected PC/IC interface are identical. The settings

for the 816 IC Eluent Selector are given in Section 4.1.2.