2 Installation

2.3 Electrical

connection

2.3.1

Operation in a modular IC system

Always switch off the 762 IC Interface and 816 IC Eluent Selector

before connecting the two instruments.


Use a 6.2134.080 or 6.2134.090 Cable to connect RS 232 interface 14

of the 816 IC Eluent Selector with a free RS 232 interface on the 762 IC

Interface according to the following diagram.

816

6.2134.080 / 6.2134.090

762

6.2134.100

PC

Figure 4:

Connecting the 816 IC Eluent Selector to 762 IC

Interface

2.3.2

Operation with a 761 Compact IC

Always switch off the 816 IC Eluent Selector and PC before connect-

ing the two instruments.

761

6.2134.100

PC

816

PC

6.2125.110

Figure 5:

Connecting the 816 IC Eluent Selector to a PC

The 816 IC Eluent Selector can also be operated with the 761 Compact

IC without a 762 IC Interface. The RS 232 interface of the two instru-

ments must then be connected directly to the RS 232 interfaces of the

PC as shown in Figure 5.

816 IC Eluent Selector/ 8.816.1003 Instructions for Use