2.2 Mains connection

2.2.2 Fuses

The fuse holder of the IC Eluent Selector is fitted with one of the follow-

ing two fuses as standard: 1.0 A/slow-blow for 100

120 V or

0.5 A/slow-blow for 220

240 V.

Never use different fuses in the instrument from those mentioned

above as otherwise there is a fire risk!

Checking and exchanging fuses is described in Section 2.2.1.

2.2.3

Mains cable and mains connection

Mains

The instrument is supplied with one of the following mains cables:

6.2122.020 with SEV 12 plug (Switzerland,

)

6.2122.040 with CEE(7), VII plug (Germany,

)

6.2133.070 with NEMA 5-15 plug (USA,

)

which has three wires and is fitted with a plug with a grounding pin. If a

different plug has to be used then the yellow/green wire (IEC standard)

must be connected to the grounding pin (protection class I).

Any interruption to the grounding inside or outside the instrument can

represent a hazard!

Mains connection

Insert the mains cable into mains connector 9 of the 816 IC Eluent Se-

lector.

2.2.4

Switching the instrument on/off

The 816 IC Eluent Selector is switched on and off with mains switch 8.

816 IC Eluent Selector/ 8.816.1003 Instructions for Use