12 detector, 13 sample degasser, 14 power connection – Metrohm 930 Compact IC Flex Oven/SeS/Deg User Manual

Page 106: 15 interfaces

Advertising
background image

7.12 Detector

■■■■■■■■■■■■■■■■■■■■■■

98

■■■■■■■■

930 Compact IC Flex Oven/SeS/Deg (2.930.2460)

7.12

Detector

You can find the technical specifications for the detector in the detector's
manual.

7.13

Sample degasser

Material

Fluoropolymer

Resistance to sol-
vents

No restriction (except PFC)

Time to establish
vacuum

< 60 s

7.14

Power connection

Required voltage

100 - 240 V ± 10% (autosensing)

Required fre-
quency

50 - 60 Hz ± 3 Hz (autosensing)

Power consump-
tion

65 W for typical analysis application

25 W standby (conductivity detector to 40 °C)

Power supply unit

Up to 300 W maximum, electronically monitored

Internal fuse 3.15 A

7.15

Interfaces

USB

Input

1 USB upstream, type B (for connection to the PC)

Output

2 USB downstream, type A

MSB

2 MSB Mini DIN 8-pin (female) (for Dosino, stirrer, remote lines, etc.)

Detector

1 15-pin high density DSUB (female)

Column recogni-
tion

for an intelligent column

Leak sensor

1 jack plug

Further connec-
tions

Auxiliary

1 15-pin DSUB (female)

Service

1 15-pin DSUB (female)