2 connecting the mcs, Connecting the mcs, Figure 14 – Metrohm 930 Compact IC Flex Oven/SeS/Deg User Manual
Page 45
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
930 Compact IC Flex Oven/SeS/Deg (2.930.2460)
■■■■■■■■
37
3.13.2
Connecting the MCS
The MCS is connected between the suppressor and the detector.
Connecting the MCS
Figure 14
Connecting the MCS
1
Air aspiration capillary
For drawing in air with low CO
2
content (via
the CO
2
adsorption cartridge).
2
Pressure screw, short (6.2744.070)
Installed on the air aspiration capillary.
3
Luer coupling (6.2744.120)
Mounted on the air aspiration capillary with
a pressure screw (6.2744.070).
1 Removing the threaded stoppers
Remove and keep the two threaded stoppers (6.2744.220) from the
input and output of the MCS.
2 Connection from the suppressor
Use a long pressure screw (6.2744.090) to connect the capillary
labeled out to the input of the of the MCS (labeled In).
3 Connection to the detector
Connect the detector input capillary with a long pressure screw
(6.2744.090) to the MCS output (labeled Out).