Metrohm NIRS XDS MultiVial Analyzer User Manual

Page 68

Advertising
background image

66

▪▪▪▪▪▪▪

Vision performs the same
operations as done on the
previous Reference
Standardization.

Click “OK”.

The Certified 80%
Reflectance Standard is
shown with the spectrum of
the ceramic instrument
reference.

Click “OK” to plot a
correction spectrum in the
next window.

The correction spectrum
represents the amount of
spectral correction required
to provide a virtual 100%
reflectance reference at each
data point.

Click “OK” to plot the
correction spectrum in the
next window.

Advertising