4.3 Maintenance and servicing

871 Advanced Bioscan / Instructions for Use 8.871.1003

25

4.3 Maintenance

and

servicing

information

4.3.1 General

truments could possibly lead to malfunc-

tions and a shorter service life of the mechanical and electronic parts.

als and solvents should be wiped up immediately. It is

Care
The 871 Advanced Bioscan requires proper care and attention. Exces-

sive contamination of the ins

Spilled chemic

especially important to protect the plug connections at the rear of the

instrument (particular the mains plug) against contamination.

Although constructional measures have been designed to

virtually

eliminate such a situation, should corrosive media penetrate the inte-

rior of the instruments the mains plug must be immediately discon-

nected to prevent extensive damage to the instrument electronics.

Inform Metrohm service if your instrument(s) have been damaged in

such a way.

The instrument must not be opened by untrained personnel. Please

comply with the safety notes in section 1.4.1.

ompartment of the 871 Advanced Bioscan a lithium

In the electronic c

battery is mounted. If this unit or battery needs service, dispose it ac-

cording to chemical waste only.

Maintenance by Metrohm service
M

p

-

q

osive chemicals, it may be nec-

essary to shorten the interval between servicing.

department is always willing to offer expert advice

on the maintenance and servicing of all Metrohm instruments.

4.3.2 Shutdow

aintenance of the IC system is best done as part of an annual service

erformed by specialists from the Metrohm company. If work is fre

uently performed with caustic and corr

The Metrohm service

n

If the IC System is shut down for a considerable length of time, the en-

tire IC system (without column) must be rinsed with methanol/water

(1:4) to avoid crystallization of eluent with the corresponding subse-

quent damage.

arating column module is removed and bridged. The

For rinsing the sep

flow cell must be turned off Rinse with methanol/water (1:4) until the

conductivity of the eluent drops below 10

µ

S/cm.