4.3.3

Cleaning the working electrode

The surface of the gold working electrode should be cleaned whenever

you notice that its electrochemical properties have changed. This may

be caused by the deposition of reaction products. High current flows

can also alter the electrode surface; this is indicated by a very reduced

sensitivity after longer periods of use.
In principle, you should avoid too high analyte concentrations or dirty

samples. This measure as well as the regenerative effect of the pulse

technique normally supersedes the polishing of the electrode.

You should on

flow

ly polish the electrode surface with the polishing paste

supplied, if the sensitivity cannot be restored by intensive rinsing of the

cell with water and cleaning the electrode surface by rubbing it

with an eraser.

Use the diamond paste (6.2802.110) supplied to polish the surface of

the electrode together with the polishing disk (6.2802.100). Proceed as

follows:

1 Rinse the polishing disk

Rinse the polishing disk with high purity water before use.

2 Shake the diamond paste

Shake the bottle containing the diamond paste thoroughly be-

fore use.

3 Polish the electrode

Add a few drops of the paste to the wet polishing disk and pol-

inute with uniform move-

form

ish the electrode for approx. one m

ments in the

of a figure 8. Exert only a slight pressure.

rode

C

condition visually and repeat the polishing

ry.

4 Clean the elect

Clean the electrode with a paper tissue moistened with ethanol.

heck the surface

process if necessa

5 Rinse electrode and measuring cell

Rinse the electrode and the measuring cell with high purity wa-

ter.

6 Mount the measuring cell

Remount the measuring cell.

7 Clean the polishing disk

Clean the polishing disk with high purity water and store it dust-

free in a plastic cover.

4.3.4

Changin

If you wish to replace an IC separating column by a column of the same

type, proceed as follows (see Figure 3):

g separating columns

Identical separation system