5.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
90
■■■■■■■■
850 Professional IC –Cation – Prep 1
Problem
Cause
Remedy
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 59).
Eluent – Gas bubbles in the
eluent.
Check connections of the eluent degasser (see
Chapter 2.9, page 33).
High pressure pump –
defective.
Request Metrohm Service.
Vacuum is not being
built
Eluent Degasser – Connec-
tor Vacuum on the rear of
the instrument not (tightly)
sealed.
■
Seal the connector Vacuum tightly with a
6.144.6040 threaded stopper.
or when using an 872 Extension Module:
■
Connect an FEP tubing between the IC
instrument and the Extension Module and
tighten both clamping screws until they
seal.