850 Professional IC –Cation – Prep 1

■■■■■■■■

31

4 Installing tubing weighting and aspiration filter

1

2

3

4

5

Figure 14

Installing tubing weighting and aspiration filter

1

Eluent aspiration tubing 6.1834.080

2

Eluent bottle attachment 6.1602.160

3

Tubing weighting
From accessory set 6.2744.210.

4

Clamping screw
From accessory set 6.2744.210.

5

Aspiration filter 6.2821.090
With filter holder from accessory set
6.2744.210.

Slide the tubing weighting (14-3) onto the eluent aspiration tub-
ing (14-1).

Slide the clamping screw (14-4) onto the eluent aspiration tubing
(14-1).

Insert eluent aspiration tubing (14-1) into the aspiration filter
(14-5). The end of the tubing must touch the base of the filter.

Screw together clamping screw (14-4) and filter holder (13-1).
The end of the tubing must still touch the base of the filter.

Figure 15

Eluent aspiration tubing fully equipped.

5 Mounting eluent aspiration tubing to the eluent bottle

Insert the eluent aspiration tubing into the eluent bottle (16-10).