2 electrical safety, Electrical safety – Metrohm 875 KF Gas Analyzer User Manual

Page 13

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

1 Introduction

■■■■■■■■

5

NOTE

Check all connections of the system for leakage at regular intervals and
particularly after having made any modifications.

WARNING

The gas system is under pressure. It contains both pressurized gases
and liquefied gases.

Before the sample vessel can be changed, the pressure must be
released in the piping system and the latter may need to be rinsed with
nitrogen.

Observe the applicable regulations.

WARNING

The oven used for vaporizing the liquefied gases and the oil filter down-
stream of the oven may exhibit a temperature of up to 70 °C. Avoid
direct skin contact. Wear heat-insulating gloves, if necessary.

Clean the oil filter and rinse the piping carrying gas through the oven
only with the instrument switched off and while it is cold.

1.5.2

Electrical safety

The electrical safety when working with the instrument is ensured as part
of the international standard IEC 61010.

WARNING

Only personnel qualified by Metrohm are authorized to carry out service
work on electronic components.

WARNING

Never open the housing of the instrument. The instrument could be
damaged by this. There is also a risk of serious injury if live components
are touched.

There are no parts inside the housing which can be serviced or replaced
by the user.

Advertising