Metrohm 875 KF Gas Analyzer User Manual

Page 27

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

■■■■■■■■

19

Partial step

Condition

Opened valves

Stop condition

Draining of the sam-
ple that flowed into
the area upstream of
the regulator

Method variable "first
sample measure-
ment?" is set to "yes"

Waste gas valve

60 seconds expired

Rinsing out the waste
gas line with nitrogen

Method variable "first
sample measure-
ment?" is set to "yes"

Nitrogen valve

Waste gas valve

45 seconds expired

Prerinsing with nitro-
gen

None

Nitrogen valve

Measurement valve

Status message from
the coulometer "Con-
ditioning OK", but at
least 60 seconds

Pressure release nitro-
gen

None

Measurement valve

20 seconds expired

Sample introduction

None

Sample input valve

Measurement valve

Value entered for
minimum sample
amount (mg) in the
method variable is
achieved

Pressure release sam-
ple

None

Measurement valve

Gas flow falls below
30 mL/min for more
than 6 seconds

Postrinsing with nitro-
gen

None

Nitrogen valve

Measurement valve

Stop criteria of the
coulometric KF titra-
tion are met (extrac-
tion time and relative
drift)

Relieving the sample
infeed

Method variable "dis-
connect gas container
after measurement?"
is set to "yes"

Sample valve

Stopcock 1