background image

4.2 Methods

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

7 Check whether a valid value is entered under the corresponding

common variable in the configuration.

8 Load the Sample measurement method in the tiamo™ Methods

part with File

Open and double-click on the A6 calculation com-

mand to open it.

9 Edit the line A7 checking gas volume via the properties and open

the formula editor by clicking on the ÷ symbol. An if-then query (nes-
ted CASE function) then opens; in this query the additional gas type x
that you have replaced with butadiene is listed twice in a row.
Replace the term additional gas type x with butadiene also here and
save the method using File

Save. You can now select your new

gas type for the subsequent analyses, and the method automatically
uses the appropriate correction factor for the calculations.