2 working steps for carrying out a measurement, Working steps for carrying out a measurement, Figure 6 – Metrohm 875 KF Gas Analyzer User Manual
Page 28
Advertising
4.2 Methods
■■■■■■■■■■■■■■■■■■■■■■
20
■■■■■■■■
Figure 6
Schematic representation of the gas flows during an analysis
Red marking = sample flow
Green marking = nitrogen
1
Prerinsing with sample
2
Rinsing out the waste gas line with
nitrogen
3
Prerinsing and postrinsing with nitro-
gen
4
Sample introduction
5
Relieving the sample infeed
6
Rinsing the feed line with nitrogen
4.2.2
Working steps for carrying out a measurement
Load the sample table Standard sample table gas measurement in
the run window of your tiamo™ workplace under Determination
series
▶ Sample table ▶ Load. This sample table is preset in such a way
that you can make the entries that are relevant for you. The input window
opens by double-clicking in the first line of the table template.