2 working steps for carrying out a measurement, Working steps for carrying out a measurement, Figure 6 – Metrohm 875 KF Gas Analyzer User Manual

Page 28

Advertising
background image

4.2 Methods

■■■■■■■■■■■■■■■■■■■■■■

20

■■■■■■■■

Figure 6

Schematic representation of the gas flows during an analysis

Red marking = sample flow

Green marking = nitrogen

1

Prerinsing with sample

2

Rinsing out the waste gas line with
nitrogen

3

Prerinsing and postrinsing with nitro-
gen

4

Sample introduction

5

Relieving the sample infeed

6

Rinsing the feed line with nitrogen

4.2.2

Working steps for carrying out a measurement

Load the sample table Standard sample table gas measurement in
the run window of your tiamo™ workplace under Determination
series

Sample table Load. This sample table is preset in such a way

that you can make the entries that are relevant for you. The input window
opens by double-clicking in the first line of the table template.