6 Troubleshooting
■■■■■■■■
37
yes
no
no
no
Drift > 6 µg/min
Drift diagnosis
method
→
Drift rise as a result of
rinsing > 3 µg/min
Exchange
catholyte
Dryness level of
rinsing gas?
Replace molecular
sieve cartridge
Drift < 6 µg/min?
Drift < 6 µg/min?
Rinse gas-
carrying system
Exchange
anolyte
Contact Metrohm
Service
End
Figure 9
Systematic procedure for identifying the cause of drift rises