1 sequence of the "sample measurement" method, Sequence of the "sample measurement" method – Metrohm 875 KF Gas Analyzer User Manual

Page 26

Advertising
background image

4.2 Methods

■■■■■■■■■■■■■■■■■■■■■■

18

■■■■■■■■

Gas calibration_gas

Shut down system

Drift diagnosis

System preparation

The following methods form part of the optional scope of delivery:

Rinsing with solvent

Reagent replacement

Addition of methanol

NOTE

Please note:

The tiamo™ method can only be run if the Flow program has been
started.

4.2.1

Sequence of the "Sample measurement" method

The water content determination of the samples is controlled by the Sam-
ple measurement
method, which basically consists of three steps:

Prerinsing the line route with nitrogen

Feeding in the sample

Postrinsing with nitrogen

The method is designed in such a way that the pressure prevailing in the
area before the regulator (line volume between precision control, nitrogen
and sample valve) is released during the change from prerinsing to sample
introduction and from sample introduction to postrinsing. In this way, a
mixing of nitrogen and sample that could result in faulty measurements is
prevented. The entire sequence is shown in (see Table 4, page 18).

The flow diagrams of the analysis are visualized in figure 6. Some partial
steps are only run through if the corresponding scans are set to "yes" in
the sample table. The dosing device for methanol addition and reagent
replacement as well as for the automated rinsing with solvent is an
optional equipment of the 875 KF Gas Analyzer.

Table 4

Gas conveyance and valve control during the analysis

Partial step

Condition

Opened valves

Stop condition

Prerinsing with sam-
ple

Method variable "first
sample measure-
ment?" is set to "yes"

Sample valve

Waste gas valve

90 seconds expired

Advertising