3 installation, 1 about this chapter, About this chapter – Metrohm 872 Extension Module Sample Prep User Manual

Page 14

Advertising
background image

3.1 About this chapter

■■■■■■■■■■■■■■■■■■■■■■

6

■■■■■■■■

872 Extension Modul – Sample Prep

3 Installation

3.1

About this chapter

The Installation chapter contains

this overview

a general description of mounting and connection of IC instrument
and extension module.

a brief set of instructions for the installation of the 872 Extension
Modul Sample Prep (see Chapter 3.3, page 7). At each step you will
find cross-references to more detailed installation instructions for indi-
vidual components, should you should require such aids.

an installation diagram (see Chapter 3.4, page 8) which displays an
example for a possible installation.

several chapters with extensive installation instructions for all compo-
nents.

3.2

Extension Module and 850 Professional IC instru-
ment

The extension modules are fitted directly to the 850 Professional IC instru-
ment and connected with it via 6.2156.060 connection cable. Extension
modules have no power supply of their own, but rather draw the electric-
ity they require from the instrument with which they are connected.

Extension modules can be mounted above (between instrument and cov-
ering plate) or below (between instrument and base tray) the 850 Profes-
sional IC instrument or set up next to the instrument with separate (to be
ordered separately) 6.2061.110 base tray 6.2061.100 and covering plate .

Up to three extension modules can be connected to a 850 Professional IC
instrument. The following limitation is to be taken into account when
doing so: the 850 Professional IC instrument and its extension modules
may not contain more than 4 identical components, i.e. a maximum of 4
high pressure pumps, a maximum of 4 peristaltic pumps, a maximum of 4
injection valves, a maximum of 4 suppressors (MSM, incl. SPM), but only a
maximum of 3 degassers and a maximum of 3 CO

2

suppressors (MCS).

Note

If all 4 high pressure pumps are being used at once, then not all of
them are permitted to run at maximum flow for long periods.