3 Installation

872 Extension Modul – Sample Prep

■■■■■■■■

11

3.5.3

Mounting the extension module on IC instrument

Note

Before you begin make sure that the preparation work (see Chapter
3.5.1, page 9)
has all been carried out.

Dismounting covering plate

2

2

1

1

Figure 5

Covering plate from above

1

Cylinder screws

2

Covering stoppers

1 Dismounting covering plate

Remove covering stoppers (5-2).

Undo the cylinder screws (5-1) with a 6.2621.100 3 mm hexagon
key.

Remove covering plate.

2 Mounting extension module

Place the extension module onto the IC instrument.