2 installing the peristaltic pump, Installing the peristaltic pump, Figure 7 – Metrohm 872 Extension Module Sample Prep User Manual
Page 22: Peristaltic pump, 3) and the tub
Advertising
3.7 Peristaltic pump
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
872 Extension Modul – Sample Prep
6
1
4
3
2
5
7
Figure 7
Peristaltic pump
1
Knurled screw in the mounting pin
2
Roller hub
3
Rollers
4
Cartridge holder
5
Tubing cartridges 6.2755.000
6
Contact pressure lever
7
Snap-action lever
3.7.2
Installing the peristaltic pump
Note
Depending on the application of the peristaltic pump, either a
6.2744.180 pump tubing connection with filter or a 6.2744.160 pump
tubing connection without filter can be used on the pressure side.
For pumping the auxiliary solutions to the MSM or to the SPM, a
6.2744.180 pump tubing connection with filter must be used.