6.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

872 Extension Modul – Sample Prep

6 Troubleshooting

6.1

Problems and their solutions

Problem

Cause

Remedy

Peristaltic pump –
insufficient or no
delivery rate

Peristaltic pump – contact
pressure too weak.

Correctly set contact pressure (see "Setting the
flow rate", page 18)
.

Peristaltic pump – filter
blocked.

Replacing filter (see Chapter 5.2.2.2, page 23).

Peristaltic pump – pump
tubing defective.

Replace pump tubing (see Chapter 5.2.2.1,
page 22)
.