5 Handling and maintenance

872 Extension Modul – Sample Prep

■■■■■■■■

23

Order
number

Name

Material

Inner
diameter

Use

6.1826.360

Pump tubing LFL
(white/white), 3-stop-
per

PVC (Tygon)

1.02 mm

For sample transfer.

6.1826.380

Pump tubing LFL
(gray/gray), 3-stopper

PVC (Tygon)

1.25 mm

For inline sample dilution.

6.1826.390

Pump tubing LFL (yel-
low/yellow), 3-stop-
per

PVC (Tygon)

1.37 mm

For the sample solution in
inline ultrafiltration.

5.2.2.2

Pump tubing connector with filter

The 6.2821.130 filters (13-2) should be changed every 3 months, more
frequently at higher backpressure.

1

2

3

Figure 13

Pump tubing connection – changing the filter

1

Tubing olive

2

Filter 6.2821.130
Packaging contains 10 items.

3

Filter screw

Replacing the filter

1 Unscrewing the filter screw

Unscrew the filter screw (13-3) out of the tubing olive (13-1).

2 Replacing the filter

Remove old filter (13-2).

Place new filter (13-2) flat in the tubing olive (13-1).

3 Mounting the filter screw

Screw filter screw (13-3) back into the tubing olive (13-1).