Application Bulletin 403/1 e

Installation instruction for MVA-21

4.2.4. Results

In the «Results» part on the tab «Results» the substance is

selected for which the concentration should be calculated

and displayed.

5. Operation

5.1. Suppressor determination series

5.1.1. Manual operation

Dosing Unit «894/D1 50mL VMS» connected to Dosino

1 has to be prepared with VMS.

Suppressor standard solutions and plating bath

samples are placed on position 57 to 111 of the sample

rack using the sample vials for 11 mL (6.2743.057).

On the position after the last sample a vial with pure

water is placed to rinse Dosino 4 after the end of the

suppressor sample series.

Fig 13: Example positions of solutions on the sample rack
6.2041.450 for suppressor determination

5.1.2. Workplace

In the «Run» window on the tab «Determination series»

create a sample table

Fig 14: Example sample table for suppressor determination

To run a calibration with a standard solution select

sample type S

TANDARD

. For a determination the sample

type S

AMPLE

has to be selected.

No «Sample amount» needs to be defined in the

sample table, since addition volumes for the standard

and the sample are defined in the method

commands ADD

STD and ADD

SAMPLE

DT.

5.1.3. Course of events

The execution of calibration and determination by

dilution titration is controlled by the 894 Professional

CVS and viva.

a) The measuring vessel is emptied using the 843

Pump Station.

b) Dosing Unit «894/D4 2mL Std or sample»

connected to Dosino 4 is automatically rinsed and

prepared with suppressor standard or sample from

the vial on the rack of the sample changer.

c) The measuring vessel is rinsed using the 843

Pump Station.

d) Dosino 1 automatically adds the VMS and Dosino

4 is used for the automatic addition of suppressor

standard or plating bath sample.

e) At the end of each determination the measuring

vessel, is automatically emptied and rinsed using

the 843 Pump Station.

Next sample starts again from a).

Recalibration within a determination series is possible.

At the end of a determination series Dosino 4 is

automatically prepared with rinsing solution.