Application Bulletin 403/1 e

Installation instruction for MVA-21

6. Remarks

6.1.

Single additive setup

If the system is only used for suppressor or brightener

determination the unused port at the T connector

(6.2744.290) has to be closed with a threaded stopper

(6.2744.060).

6.2. Combination of applications

The determination of suppressor by dilution titration (DT)

and the determination of brightener by modified linear

approximation technique (MLAT) can also be combined in

one series.

6.3.

Storing dosing units

When the dosing units are not used (during the night, over

the weekend) the dosing cylinder has to be rinsed with

water. Otherwise elemental copper can be formed or

additives can precipitate between the valve disk and the

distributor disk or in the tubings. A blockage can damage

the 800 Dosino. Emptying the cylinder and filling it with

water is sufficient to prevent a blockage.

6.4.

Storing electrodes

When the electrodes are not used (during the night, over the

weekend) the electrodes should be thoroughly rinsed.

Working and auxiliary electrode can either be stored in

deionized water or dry. The reference electrode should be

stored separately in a vial filled with deionized water (KCl

solution, in case KCl is used as bridge electrolyte), so that

the reference electrode immerses at least to the rim of the

electrolyte vessel.

6.5.

Waste container

The waste container must not be closed completely. For

pressure balancing in the waste container, keep at least one

opening unsealed. Overpressure would lead to a

malfunction of the drain pump.

Page 12 of 12