4.1.5. Rinsing measuring vessel

After each sample the measuring vessel need to be rinsed

with deionized water. The rinsing and draining times depend

on the total volume used in the determination.

If the method templates are used, the rinsing and draining

times are defined in the corresponding PUMP

commands in

the TRACK

–R

INSE

and

TRACK

D

RAIN

. The number of

rinsing cycles is defined in the command LOOP

LOOP

R

INSING

in the TRACK

R

INSING CYCLES

.

Fig 11: Method snippet TRACK

R

INSING CYCLE

Examples for rinsing and draining times:

Cell volume

50 mL

100 mL

Number of rinsing cycles

2

2

Draining time [s]

35

50

Rinsing time [s]

9

18

Please note! The device assigned to the commands PUMP

PUMP

R

INSE

und PUPM

PUMP

D

RAIN

need to be the

858 Professional Sample Processor, since the 843 Pump

Station is connected to and controlled from the sample

changer.

4.1.6. Rinsing transfer tubing

This only applies for the brightener determination by MLAT!

After each sample the transfer tubing from the 858

Professional Sample Processor to the measuring vessel of

the 894 Professional CVS need to be rinsed with deionized

water. The rinsing solution is placed in a vial on the rack of

the 858 Professional Sample Processor. The position of the

rinsing solution relative to the sample position is defined in

the

TRACK

G

O TO RINSING POSITION

in the

command MOVE

MOVE

R

INSING POSITION

.

Fig 12: Method snippet TRACK

G

O TO RINSING POSITION

The position of the rinsing solution is defined as «sample

position + 28». This means the outer circle on the rack

6.2041.450 is used for the sample the inner circle for the

rinsing solution

The transfer time for the rinsing solution corresponds to the

sample transfer time defined in the TRACK

P

ERISTALTIC

PUMP

in the

command PUMP

P

ERISTALTIC PUMP

ON.

4.2.

Evaluation

Settings regarding evaluation and documentation of the

determination are located in the «Evaluation» part of the

method. The templates already include all necessary

settings to determine brightener or suppressor concentration

in an acid copper bath. If modifications however should be

necessary, here is where important parameters are found:

4.2.1. Substances

In the «Substances» part settings for peak recognition and

baseline parameters are defined.

4.2.2. Standards

In the «Standards» part the concentration of the used

standard solution is defined.

4.2.3. Calibration

In the «Calibration» part the calibration method, such as DT

or MLAT, is defined as well as the regression type.