Application Bulletin 403/1 e

Installation instruction for MVA-21

Swing head

Swing position

0

mm

Rinse position

0

mm

External position

Not used

3.5.2. Rack

Rack name

6.2141.450

Shift rate

20

°/s

3.5.3. Rack data

Rack parameters

Beaker radius samples

off

mm

Beaker sensor

off

Rack offset

0

°

Lift positions – Tower 1

Work position

User

defined

mm

Rinse position

0

mm

Shift position

0

mm

Special position

0

mm

Lift positions – Tower 2

Work position

Not available

Rinse position

Not available

Shift position

Not available

Special position

Not available

Special beakers

Not used

3.5.4. Needle adjustment

The work position of the pipetting needle 6.1835.050 has to

be adjusted in a way that the lower end of the needle is

positioned max. 1 mm above the bottom of the sample vial.

This is essential to guarantee a complete transfer of the

sample from the sample vial on the rack of the 858

Professional Sample Processor into the measuring vessel of

the 894 Professional CVS.

Fig 9: Needle adjustment

Procedure to adjust the needle position in viva

1. Insert an empty sample vial to the desired sample

position on the rack (position 57 to 126).

2. Click on the Manual symbol.

3. Select Tower 1 (Sample changers – 858 Professional

Sample Processor).

4. Open tab «Move».

5. Enter the position of the empty sample vial in the field

«Target position».

6. Click Start to move the needle to the chosen rack

position.

7. Enter 125 mm in the field «Target position» in the

section «Lift position».

8. Click Start to lower the needle to the chosen target

position. To adjust the needle like shown above, use

the arrow up and down buttons

to lower the

needle stepwise.

0.5 … 1 mm