Dosing unit at 800 dosino (2) and 800 dosino (3), Dosing unit at 800 dosino (4), Solutions – Metrohm MVA-21 User Manual

Page 6: Tower

Advertising
background image

Application Bulletin 403/1 e

Installation instruction for MVA-21

Dosing rate port 1

Maximum

mL/L

Dosing rate port 2

Maximum

mL/L

Dosing rate Fill port

Maximum

mL/L

Dosing rate special port

Maximum

mL/L

Tubing parameters

Port

Length

Diameter

Dosing port 1

Port 1

138 cm

2 mm

Dosing port 2

Port 3

0 cm

2 mm

Fill port

Port 2

25 cm

2 mm

Special port

Port 4

0 cm

2 mm

Valve disk

Rotating direction

automatic

3.3.2. Dosing unit at 800 Dosino (2) and 800 Dosino (3)

Dosing Unit name as used in the viva method templates.

800 Dosino (MSB 2)

894/D2 2mL Brightener

800 Dosino (MSB 3)

894/D3 2mL Suppressor

Parameters for preparation

Dosing port Prep/empty

Dosing port 2

Dosing rate port 1

2.0

mL/L

Dosing rate port 2

maximum

mL/L

Dosing rate Fill port

Maximum

mL/L

Dosing rate special port

Maximum

mL/L

Tubing parameters

Port

Length

Diameter

Dosing port 1

Port 1

80 cm

0.3 mm

Dosing port 2

Port 3

0 cm

2 mm

Fill port

Port 2

55 cm

2 mm

Special port

Port 4

0 cm

2 mm

Valve disk

Rotating direction

automatic

3.3.3. Dosing unit at 800 Dosino (4)

Dosing Unit name as used in the viva method templates.

800 Dosino (MSB 4)

894/D4 2mL Std or sample

Parameters for preparation

Dosing port Prep/empty

Dosing port 2

Dosing rate port 1

2.0

mL/L

Dosing rate port 2

maximum

mL/L

Dosing rate Fill port

4.0

mL/L

Dosing rate special port

maximum

mL/L

Tubing parameters

Port

Length

Diameter

Dosing port 1

Port 1

80 cm

0.3 mm

Dosing port 2

Port 3

0 cm

2 mm

Fill port

Port 2

230 cm

1 mm

Special port

Port 4

0 cm

2 mm

Valve disk

Rotating direction

automatic

3.4.

Solutions

Solutions that should automatically be dosed have to be

defined in the viva «Configuration» and need to be

assigned to the Dosing Unit which is used for the dosing.

The following table shows the solution names and assigned

Dosing Units as used in the viva method templates.

Solution name

Dosing Unit

VMS

894/D1 50mL VMS

Brightener concentrate

894/D2 2mL Brightener

Suppressor concentrate

894/D3 2mL Suppressor

Std or sample

894/D4 2mL Std or sample

S

OLUTION TYPE

for «VMS» and «Suppressor concentrate»

has to be A

UXILIARY SOLUTION

, for «Brightener concentrate»

and «Std or sample» it is S

TANDARD SOLUTION

.

For a step by step description please see the Tutorial

chapter «5.1.4 Define solutions».

3.5.

Devices – 858 Professional Sample Processor for

CVS

The «858 Professional Sample Processor for CVS» is

automatically recognized by the viva software. When an

instrument is connected for the first time it needs to be

entered in the viva «Configuration». The corresponding

dialogue will pop up automatically.

Device name*

858 Professional Sample Processor 1

* as used in the example method

For a step by step description please see the Tutorial

chapter «6.1.1 Configuring the instrument».

3.5.1. Tower

Tower parameters

Max. stroke path

130

mm

Min. beaker radius

Off

Lift rate

25 mm/s

mm/s

Axial distance

166.0

mm

Page 6 of 12