Brightener determination series, Manual operation, Workplace – Metrohm MVA-21 User Manual

Page 11: Course of events, Remarks

Advertising
background image

Application Bulletin 403/1 e

Installation instruction for MVA-21

5.2. Brightener determination series

5.2.1. Manual operation

Dosing Unit «894/D1 50mL VMS» connected to Dosino

1 has to be prepared with VMS.

Dosing Unit «894/D2 2mL Brightener» connected to

Dosino 2 has to be prepared with brightener

concentrate.

Dosing Unit «894/D3 2mL Suppressor» connected to

Dosino 3 has to be prepared with suppressor

concentrate.

Plating bath samples are placed on position 1 to 28 of

the sample rack using the sample vials for 50 mL

(6.2747.010). ). The exact sample volume needed for

the determination has to be pipetted into the vial.

Rinsing solutions are placed on position 29 to 56 of the

sample rack using the sample vials for 50 mL

(6.2747.010). Usually the same volume of distilled

water is used for rinsing like it was used for the sample.

The rinsing solution has to be placed on the inner circle

on the neighboring position of the sample (Sample

position + 28).

Fig 15: Example positions of solutions on the sample rack
6.2041.450

5.2.2. Workplace

In the «Run» window on the tab «Determination series»

create a sample table

Fig 16: Example sample table

As sample type S

AMPLE

has to be selected.

For «Sample amount» the volume pipetted into the vial

has to be entered.

5.2.3. Course of events

The execution the determination by modified linear

approximation technique is controlled by the 894

Professional CVS and viva.

a) The measuring vessel is emptied using the 843

Pump Station.

b) The intercept solution is automatically prepared at

the beginning of each determination, by dosing

VMS from Dosino 1 and suppressor concentrate

from Dosino 3.

c) After the determination of the intercept value, the

plating bath sample is transferred from the sample

rack to the measuring vessel by means of the

build-in peristaltic pump of the 858 Professional

Sample Processor. The sample is added on top of

the intercept solution.

d) Standard addition is carried out automatically with

brightener concentrate from Dosino 2.

e) At the end of each determination the measuring

vessel is emptied using the 843 Pump Station.

f)

The transfer tubing from the sample changer to the

measuring vessel is rinsed with the rinsing solution

placed on the rack of the sample changer.

g) After the transfer tubing is rinsed the measuring

vessel is automatically emptied and rinsed using

the 843 Pump Station.

Next sample starts again from a).

Page 11 of 12