3 configuring the rs-485 port – HF scientific AccUView Wastewater UV %Transmission Monitor User Manual

Page 24

Advertising
background image

The next, prompt will be the level assigned to the 20 mA output level (

UPLM on the

lower row of the LCD display). Select the %T level to assign to the

UPLM using the

t

and

u buttons. Once the desired level has been set, press the ↵ button to accept it.











The final prompt for the 4-20 mA setup is for the error level (

ERLV). This is the current

level that the instrument will output when an error occurs for possible errors see section
10.1. The error level selections are 0.00 mA, 2.00 mA, 4.00 mA , or OFF to disable this
feature. The default setting is 2.00 mA.









7.3 Configuring the RS–485 Port
If the instrument is equipped with this option (HF cat # 19851A), and the I/O selection is
changed to

485, prompts will appear for setting the baud rate and the address.


Select the correct baud rate (1200, 2400, 4800, 9600, or 19200) for operation of the I/O
port by pressing the

t or u buttons to change the displayed baud rate.

Press the

button to continue on and select the desired instrument address using the t or

u buttons. Once the selection is made, press the ↵ button to save the setting.

AccUView Wastewater (1/09)