7.12 Pumping Time
This is amount of time in minutes the pump will operate in each pump cycle. The default
is set to 5 minutes. If no changes are made to pump settings the pump will run for 5
minutes every 30 minutes










7.13 Settling Time
This is the period of time in seconds where the sample is allowed to settle or degas prior to
measuring. The default setting is 20 seconds and adjustment is allowed from 1 to 60
seconds.











7.14 Measurement Time
This portion of the cycle is where the actual measurements are taken. Measurements are
taken during the lamp on time (once every 10 seconds). The default setting of 60 seconds
allows for 6 consecutive readings all of which are added to the averaging buffer.
Allowable adjustment is from 20-150 seconds.