Specifications – HF scientific AccUView Wastewater UV %Transmission Monitor User Manual

Page 7

Advertising
background image

Specifications

Measurement Range

0 – 100.0 %T

Reproducibility

± 0.1 %T

Resolution

0.1 %T

Accuracy

± 1.0 %T (30%T to 100%T)

Pump Draw

3.6 meters (12 feet)

Flow Rate

0.5 – 1.0 liter/min depending on lift required

Path Length

22 mm(Readings corrected to 10 mm)

Light Source

Low Pressure Mercury Lamp – 254nm

Display

Multi-Line Liquid Crystal Backlit Display

Alarm

Programmable, 120-240VAC 2A Form C Relay

Analog Output

Powered loop isolated 4-20 mA, 600

Ω drive

Communications Port

Optional Bi-directional RS-485 with Modbus

Maximum Water Pressure

1380 kPa (200 psi.) with integral pressure regulator

Operating Temperature

0°C – 50°C (32°F – 122°F)

Model 19571C: -10°C – 50°C (14°F – 122°F)

Wetted Materials

Nylon, Quartz, Silicon, Polypropylene, Stainless Steel

Sample Temperature Range

0°C – 50°C (34°F – 122°F)

Power Supply

100– 240 VAC, 47 – 63 Hz, 80VA

Insulation Rating

Double Insulated, Pollution Degree 2, Overvoltage Category II

Environmental Conditions

Not recommended for outdoor use.

Altitude up to 2000 meters

Up to 95 % RH (non-condensing)

Enclosure Rating

Designed to meet IP 66 /NEMA 4X

Regulatory Compliance
And Certifications

CE Approved, ETL listed to UL 61010B-1 &

ETL Certified to CSA 22.2 No. 1010-1-92

Shipping Weight

23.6 kg (52 lbs.)

Warranty

1 Year from date of shipment