10.4 Diagnostic Chart

Symptom

Cause Cure

Lower display shows

MA

4-20 mA loop open

Check wiring. See sections 3.3.4
and 7.2

Lower display shows

DESC

Desiccant

pouch

bad

Change desiccant pouch. See
section 9.2

Lower display shows

LAMP

Lamp failed

Replace lamp. Refer to section 9.3

Lower display shows

CLN

Ultrasonic cleaning failure

Refer to section 8.1

Lower display shows

CAL Err

after a calibration.

100%T Calibration failure

Refer to section 5.2

Lower display shows

FAIL

Major system fault

Refer to section 10.1 & 10.2

Readings are lower than expected

(1) Bubbles in solution

(2) Condensate or leaky

cuvette


(3) Flow through cuvette

dirty

Instrument out of calibration

Increase P-ST. Refer to section 7.13

Check flow through cuvette for

condensate or leaks.


Clean cuvette. See section 9.1


Recalibrate. Refer to section 5

Readings are erratic

(1) Bubbles in solution
(2) Debris in flow through

(1) See above
(2) Clean debris from cuvette

Readings are higher than expected

Instrument out of calibration

Recalibrate. Refer to section 5


10.5 Technical and Customer Assistance
If for any reason assistance is needed regarding this instrument please do not hesitate to
contact either the HF scientific Technical Service Department or the HF scientific
Customer Service Department. Contact information is shown on section 10.2.